ENDO LINEAR CUTTERS
Report
- Report Number
- 3005075853-2016-07053
- Event Type
- Injury
- Date Received
- December 15, 2016
- Date of Event
- November 17, 2016
- Report Date
- November 28, 2016
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4) DATE SENT: 1/12/2017. ADDITIONAL INFORMATION REQUESTED AND RECEIVED: WHAT WERE THE INDICATIONS FOR SURGERY? - BARIATRIC SURGERY (GASTROPLASTY) TECHNIQUE BYPASS. HOW WAS THE DEFECT/FISTULA CORRECTED IN PRIMARY PROCEDURE? - THE PROBLEM WAS CORRECTED WITH THE USE OF SUTURE WIRE, WHOLE GASTRO ANASTOMOSIS. HOW LONG POST-OP DID ISSUE OCCUR? - 04 DAYS AFTER PROCEDURE WHAT COLOR CARTRIDGES WERE USED? - BLUE CARTRIDGE 6R45B 05 UNITS - WHITE CARTRIDGE TR45W 02 UNITS PLEASE DESCRIBE STAPLE FORM. WAS ENTIRE STAPLE LINE MALFORMED? - THERE WAS A BAD FORMATION ON 01 CLAMP ACROSS THE LINE. IS IT ROUTINE FOR SURGEON TO PLACE DRAIN FOR THIS SURGICAL PROCEDURE? - TEAM USES OUR DRAIN IN ALL PROCEDURE CODE 2228 (15 FR) AND 2160 (RESERVOIR). WHAT IS THE CURRENT PATIENT STATUS? - PATIENT WAS DISCHARGED, AFTER RETURN OF THE HEAD OF THE TEAM WILL HAVE MORE DETAILED INFORMATION OF THE CURRENT STATE OF THE PATIENT.
(B)(4). THE LOT HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT.
IT WAS REPORTED THAT AFTER A GASTROPLASTY BYPASS PROCEDURE, THERE WAS A POOR FORMATION IN THE STAPLE LINE IN A STAPLE (FISTULA), WHICH REQUIRED FURTHER INTERVENTION. CORRECT SURGICAL PROTOCOL, BUT AFTER DISCHARGE OF PATIENT, THE PATIENT PRESENTED PAIN AND A LEAK WAS DETECTED BY THE DRAIN BY THE METHYLENE BLUE TEST. THIS LED TO A FURTHER HOSPITALIZATION AFTER FOUR DAYS FOR CORRECTION OF THE ENTIRE GASTRO ANASTOMOSIS FISTULA WITH SUTURE. ONE DEVICE WAS DISCARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 828978 | ENDO LINEAR CUTTERS | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | N4L54R |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |