FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 6175980 · Received December 15, 2016

Report

Report Number
1720753-2016-03401
Event Type
Malfunction
Date Received
December 15, 2016
Date of Event
November 21, 2016
Report Date
December 14, 2016
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
OWB
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FLUOROSCOPY FUNCTIONS (FFB) PCB, GENERATOR INTERFACE (GIB) PCB, CONTROL PANEL PROCESSOR PCB AND BACKPLANE WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
829173 9800 FLUOROSCOPIC X-RAY OWB GE OEC MEDICAL SYSTEMS (SLC) 9800 82-0309