FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 6175931
·
Received December 15, 2016
Report
- Report Number
- 1720753-2016-03400
- Event Type
- Malfunction
- Date Received
- December 15, 2016
- Date of Event
- November 20, 2016
- Report Date
- December 14, 2016
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- OWB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE INTELLIGENT SHUTDOWN PCB WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 826440 | 9900 | FLUOROSCOPIC X-RAY | OWB | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |