FDA Adverse Event Injury Summary report: N

ADAPTER NEO-VERSO INFANT/NEO Y W/DIR CON

MDR report key: 6175564 · Received December 15, 2016

Report

Report Number
3010838917-2016-00019
Event Type
Injury
Date Received
December 15, 2016
Date of Event
November 1, 2016
Report Date
April 24, 2017
Manufacturer
AOK TOOLING LIMITED
Product Code
CAI
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6) -EXP - THE COMPLAINT SAMPLE WAS PROVIDED TO THE MANUFACTURING FACILITY AND AN INVESTIGATION WAS COMPLETED. THROUGH THE STUDY OF THE APPEARANCE AND MEASUREMENTS OF THE DIMENSIONS OF THE RETURNED COMPLAINT PRODUCT, THE PRODUCT WAS FOUND TO MEET ALL DIMENSIONAL SPECIFICATIONS. THROUGH THIS INVESTIGATION, NO ISSUES WERE FOUND WITH THE COMPLAINT SAMPLE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED FOR THIS LOT. THERE WERE NO ISSUES IDENTIFIED WITH THE MATERIAL OR MANUFACTURING PROCESS THAT WOULD HAVE CONTRIBUTED TO THE REPORTED ISSUE.

Additional Manufacturer Narrative · 1

(B)(4) - THE COMPLAINT SAMPLE HAS NOT BEEN PROVIDED FOR EVALUATION AT THIS TIME. IF THE COMPLAINT SAMPLE BECOMES AVAILABLE, A FOLLOW-UP SUBMISSION WILL BE COMPLETED. (B)(4).

Description of Event or Problem · 1

UNDER INVESTIGATION COMPATIBILITY PROBLEM WITH DRAEGER BABYLOG NEONATAL Y-PIECE FLOW SENSOR. ACCORDING TO THE DISTRIBUTOR, A CUSTOMER (SALERNO) HAS STARTED TO USE NEO-VERSO WITH DRAEGER BABYLOG NEONATAL Y-PIECE FLOW SENSOR; THEY USE VENTILATOR WITH F&P HUMIDIFIER. AFTER SOME MINUTES OF USE, NEO-VERSO SLIPS OUT AND THE PATIENT STOPS TO BE VENTILATED WITH HIGH RISK TO DIE. THEY TESTED NEO-VERSO WITH MORE THAN ONE BABYLOG FLOW SENSOR (ALL NEW) BUT THE ISSUE STAYS. BEFORE USING VERSO, THE CUSTOMER USED KIMBERLY CLARKS CLOSED SUCTION SYSTEM WITHOUT ANY PROBLEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
826768 ADAPTER NEO-VERSO INFANT/NEO Y W/DIR CON CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) CAI AOK TOOLING LIMITED 201004130002

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| R