FDA Adverse Event
Other
Summary report: N
BD LOGIC BLOOD GLUCOSE METER
MDR report key: 617338
·
Received August 5, 2004
Report
- Report Number
- 2243072-2004-00448
- Event Type
- Other
- Date Received
- August 5, 2004
- Date of Event
- July 8, 2004
- Report Date
- August 4, 2004
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED WITH CONCERNS ON ACCURACY. THEY REPORT THEIR SPOUSE IS COMPARING METER TO METER RESULTS. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READING (140, 134) AS REF. POINT VS. BD READINGS (288, 206) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LOGIC BLOOD GLUCOSE METER | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 3064282 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |