FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 6173332
·
Received December 14, 2016
Report
- Report Number
- 1314492-2016-05297
- Event Type
- Malfunction
- Date Received
- December 14, 2016
- Date of Event
- October 1, 2016
- Report Date
- December 14, 2016
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYSTEM ERROR 105 WHICH WAS REPRODUCED AT POWER UP. SYSTEM ERROR 105 WAS CONFIRMED THROUGH A REVIEW OF THE EVENT HISTORY LOG AND FOUND TO BE CAUSED BY A SEIZED MOTOR. THE FAILED MOTOR ASSEMBLY WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 105. ANY PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 823567 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |