FDA Adverse Event Malfunction Summary report: N

INOMAX DSIR (DELIVERY SYSTEM)

MDR report key: 6173203 · Received December 14, 2016

Report

Report Number
3004531588-2016-00081
Event Type
Malfunction
Date Received
December 14, 2016
Report Date
December 13, 2016
Manufacturer
MALLINCKRODT MANUFACTURING LLC
Product Code
MRN
PMA / PMN Number
K061901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORTABLE MALFUNCTION WITH INOMAX DSIR (B)(4) WAS CREATED AS (B)(4). THE SERVICE LOG REVIEW REVEALED A DELIVERY FAILURE ALARM DUE TO MAIN SUPPLY VOLTAGE OUT OF TOLERANCE (VALID RANGE: 2435 TO 4075 COUNTS)(ACTUAL VALUE: 1934 COUNTS). THIS IS AN ERROR AS A LOW BATTERY ALARM SHOULD ALERT THE USER PRIOR TO THE DELIVERY FAILURE ALARM. A FULL FUNCTIONAL TEST WAS PERFORMED AND THE DEVICE OPERATED ACCORDING TO SPECIFICATIONS SO IT WAS RETURNED TO THE FIELD. THE ROOT CAUSE FOR THIS REPORT WAS SMART BATTERY FUEL GAUGE DRIFT. THIS CONDITION WILL BE TRACKED AND TRENDED UNDER THE COMPANY'S QUALITY SYSTEM. THIS CASE DID NOT RESULT IN AN ADVERSE EVENT/SERIOUS ADVERSE EVENT, HOWEVER, IT IS BEING SUBMITTED TO REGULATORY AUTHORITIES BECAUSE A SIMILAR FAILURE OCCURRED IN THE PAST WHICH RESULTED IN A SERIOUS ADVERSE EVENT (MDR 3004531588-2014-00002).

Description of Event or Problem · 1

ON (B)(6) 2016, A MALLINCKRODT INTERNAL EMPLOYEE DISCOVERED A DELIVERY FAILURE ALARM DUE TO MAIN SUPPLY VOLTAGE OUT OF TOLERANCE DURING ROUTINE SERVICE LOG REVIEW FOR INOMAX DSIR (B)(4). THIS MATCH WAS FOUND DURING ROUTINE REVIEW OF PREVENTATIVE MAINTENANCE SERVICE LOGS. THIS IS BEING REPORTED AS IT IS CONSIDERED A REPORTABLE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
825190 INOMAX DSIR (DELIVERY SYSTEM) APPARATUS MRN MALLINCKRODT MANUFACTURING LLC 10071

Patients

Seq Age Sex Outcome Treatment
1