FDA Adverse Event Malfunction Summary report: N

C-SHAPE SLV RDCTN STYLE

MDR report key: 6171945 · Received December 13, 2016

Report

Report Number
3004485144-2016-00371
Event Type
Malfunction
Date Received
December 13, 2016
Date of Event
November 14, 2016
Report Date
December 13, 2016
Manufacturer
ZIMMER BIOMET SPINE
Product Code
NKB
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WITHOUT A PRODUCT RETURN, NO PRODUCT EVALUATION IS ABLE TO BE CONDUCTED. THE LOT NUMBER IS UNKNOWN; THEREFORE THE DEVICE HISTORY RECORDS ARE UNABLE TO BE REVIEWED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT. THIS IS REPORT FOUR OF SIX FOR THIS EVENT. REFERENCE REPORTS 3004485144-2016-00367 THRU 3004485144-2016-00368 AND 3004485144-2016-003070 THRU 3004485144-2016-00373.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CLOSURE TOPS WERE STRIPPING WITHIN THE SHORT EXTENSION SLEEVES. THE CLOSURE TOPS WERE REPLACED AND THE FINAL TIGHTENING WAS PERFORMED WITH THE COUNTER TORQUE INSTEAD OF THE SLEEVES. THERE WAS NO PATIENT INJURY REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
821064 C-SHAPE SLV RDCTN STYLE PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM NKB ZIMMER BIOMET SPINE N/A UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1