V60 VENTILATOR

MDR report key: 6171240 · Received December 13, 2016

Report

Report Number
2031642-2016-03446
Event Type
Malfunction
Date Received
December 13, 2016
Report Date
November 15, 2016
Manufacturer
RESPIRONICS CALIFORNIA, INC
Product Code
MNT
PMA / PMN Number
K082660
Removal / Correction Number
Z-2061-2017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER REPLACED THE V60 DATA ACQUISITION (DA) TO MOTOR CONTROLLER (MC) CABLE/MOTOR CONTROLLER (MC) BRACKET RETROFIT KIT. UNIT PASSED THE REQUIRED TEST OF THE PERFORMANCE VERIFICATION SUCCESSFULLY.

Additional Manufacturer Narrative · 1

CORRECTED. THE FIELD SERVICE ENGINEER REPLACED THE V60 DATA ACQUISITION (DA) TO MOTOR CONTROLLER (MC) CABLE/MOTOR CONTROLLER (MC) BRACKET RETROFIT KIT. UNIT PASSED THE REQUIRED TEST OF THE PERFORMANCE VERIFICATION SUCCESSFULLY. THE DETERMINATION COULD NOT BE MADE THAT THE DEVICE FAILED TO MEET SPECIFICATIONS. THE DEVICE WAS NOT BEING USED FOR TREATMENT WHEN THE REPORTED EVENT OCCURRED, AND THERE IS A RELATIONSHIP OF THE DEVICE TO THE REPORTED PROBLEM. NO PARTS RETURNED TO FAILURE INVESTIGATION; THEREFORE, THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. IF PARTS ARE RETURNED, THE COMPLAINT WILL BE REOPENED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE UNIT HAS AN ERROR CODE MESSAGE OF DATA ACQUISITION/PRINTED CIRCUIT BOARD ASSEMBLY/ ANALOG DIGITAL CONVERTER (DA) PCBA ADC FAILED. THE UNIT WAS IN CLINICAL AT THE TIME THE REPORTED ISSUE WAS DISCOVERED; HOWEVER, THERE WAS NO HARM TO THE PATIENT.

Description of Event or Problem · 1

CUSTOMER REPORTED THE UNIT HAS AN ERROR CODE MESSAGE OF DATA ACQUISITION/PRINTED CIRCUIT BOARD ASSEMBLY/ ANALOG DIGITAL CONVERTER (DA) PCBA ADC FAILED. THERE WAS NO PATIENT INVOLVEMENT.