TOSOH AIA-900
Report
- Report Number
- 3005529799-2016-00017
- Event Type
- Injury
- Date Received
- December 13, 2016
- Date of Event
- November 14, 2016
- Report Date
- April 11, 2018
- Manufacturer
- TOSOH CORPORATION
- Product Code
- KHO
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- IL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
CONTACT OFFICE: (B)(6).US TELEPHONE: (B)(6). MANUFACTURING SITE: TOSOH CORPORATION (MANUFACTURER) SHIBA-KOEN FIRST BUILDING 3-8-2 SHIBA MINATO-KU, TOKYO 1058623, JA. DATE REC¿D BY MFR:HEALTH PROFESSIONAL. (B)(6) 2016. PMA/510K:510K: K971103. ON (B)(6) 2016, FSE ARRIVED AT THE SITE TO ADDRESS THE REPORTED EVENT. FSE REMOVED A WASH PROBE TIP FROM THE WASHINGTON. WELL, AND REPLACED THE WASH SYRINGE AND SEALS. FSE ALSO CLEANED THE DETECTOR LENS, THEN PRIMED AND RAN DAILY CHECKS. NO FURTHER ERRORS WERE NOTED. ON (B)(6) 2016, THE MANUFACTURER WAS NOTIFIED OF ERRONEOUS RESULTS THAT HAD BEEN REPORTED. THE FOLLOWING CHART LISTS THE RESULTS AND SUBSEQUENT CORRECTIONS. ACCESSION # INITIAL RESULT REPEAT RESULT ANALYTE COMMENT 162910274 233 181 CA125 RESULT CHANGED 162980224 729 61.4 CA125 RESULT CHANGED 153480472 11.2 8.8 CA27-29 RESULT CHANGED, 161440474 48.5 30.93 CA27-29, RESULT CHANGED 162240440 147 91.3 , FERRITIN RESULT CHANGED 163120121 <0.10 1.52 FREE, T4 RESULT CHANGED 162980204 <0.10 0.83 FREE, T4 RESULT CHANGED 162910301 56.92 1.29 PSA RESULT CHANGED. NO FURTHER ACTION WAS REQUIRED BY FIELD SERVICE. THE MOST PROBABLE CAUSE OF THE REPORTED EVENT WAS DUE TO FAULT/ FAILURE OF THE WASH PUMP. TOSOH BIOSCIENCE, INC. IS SUBMITTING ON BEHALF OF THE FOREIGN MANUFACTURER, TOSOH CORPORATION, PER EXEMPTION NUMBER E2017013. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE IMPORTER OR MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
ON (B)(6)2016, THE CUSTOMER CALLED TO REPORT THAT FOLLOWING A B/F PROBE MAINTENANCE PROCEDURE, THEY WERE RECEIVING HIGH FLAGS ON ALL SAMPLES AND QUALITY CONTROL (QC) WITH THEIR A1A-900 ANALYZER. ON (B)(6) 2016, FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ADDRESS THE REPORTED EVENT, WHICH RESULTED IN DISCREPANT PATIENT RESULTS. THERE WAS NO INDICATION OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING.
ON 11/14/2016 TOSOH BIOSCIENCE WAS NOTIFIED THAT FOLLOWING A BF PROBE MAINTENANCE PROCEDURE, THE CUSTOMER WAS RECEIVING FLAGS FOR HIGH (EXCEEDING THE HIGH END OF THE CURVE) RESULTS. ON THIS DATE, THERE WERE NO REPORTS OF ERRONEOUS PATIENT RESULTS DURING THE PHONE NOTIFICATION. TOSOH BIOSCIENCE FIELD SERVICE ENGINEER WAS DISPATCHED AND FOUND A WASH PROBE TIP IN THE WASHING WELL. THE WASH SYRINGE SEALS WERE ALSO DETERMINED TO NEED REPLACED. REPAIRS WERE COMPLETED AND THE ANALYZER WAS WORKING AS EXPECTED. ON 11/21/2016 TOSOH BIOSCIENCE WAS NOTIFIED THAT ERRONEOUS RESULTS HAD BEEN REPORTED. (B)(6). ROOT CAUSE: WASH FUNCTION WAS NOT WORKING AS REQUIRED BECAUSE WASH PROBE TIP HAD FALLEN OFF AND SYRINGE SEALS WERE WORN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 822545 | TOSOH AIA-900 | A1A-900 | KHO | TOSOH CORPORATION | AIA-900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |