PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 6170666 · Received December 13, 2016

Report

Report Number
2032227-2016-49525
Event Type
Malfunction
Date Received
December 13, 2016
Date of Event
November 28, 2016
Report Date
December 11, 2018
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED DURING TEST. MOTOR PASSED MOTOR TEST. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT CUSTOMER RECEIVED A MOTOR ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 18.4 MMOL/L. DURING TROUBLESHOOTING, IT WAS FOUND THAT CUSTOMER ALSO RECEIVED A MOTOR POSITION ENCODER ERROR ALARM. CUSTOMER REPORTED THAT DRIVE SUPPORT CAP APPEARED NORMAL. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
820108 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554CAH

Patients

Seq Age Sex Outcome Treatment
1