PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-49525
- Event Type
- Malfunction
- Date Received
- December 13, 2016
- Date of Event
- November 28, 2016
- Report Date
- December 11, 2018
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED DURING TEST. MOTOR PASSED MOTOR TEST. (B)(4).
IT WAS REPORTED THAT CUSTOMER RECEIVED A MOTOR ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 18.4 MMOL/L. DURING TROUBLESHOOTING, IT WAS FOUND THAT CUSTOMER ALSO RECEIVED A MOTOR POSITION ENCODER ERROR ALARM. CUSTOMER REPORTED THAT DRIVE SUPPORT CAP APPEARED NORMAL. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED.