FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS LEAD MRI SURESCAN

MDR report key: 6164031 · Received December 9, 2016

Report

Report Number
2649622-2016-15914
Event Type
Injury
Date Received
December 9, 2016
Date of Event
October 31, 2016
Report Date
June 30, 2022
Manufacturer
MPRI
Product Code
DTB
UDI-DI
00681490124812
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

IT WAS FURTHER REPORTED THAT THE RV LEAD EXHIBITED HIGH IMPEDANCE. THE RIGHT ATRIAL (RA) LEAD EXHIBITED CONFIRMED FRACTURE, UNDERSENSING AND HIGH IMPEDANCE. THE RA LEAD WAS REPROGRAMMED, FURTHERMORE THE RV LEAD AND THE RA LEAD WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED LIGHTHEADEDNESS AND TINGLING UNDER THEIR ARM. IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN IMPEDANCE ISSUE AND NON CAPTURE. A FRACTURE WAS CONFIRMED AND THE LEAD WAS PROGRAMMED OFF. THE LEAD REMAINS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
808315 CAPSUREFIX NOVUS LEAD MRI SURESCAN PERMANENT PACEMAKER ELECTRODE DTB MPRI 5076-52 00681490124812

Patients

Seq Age Sex Outcome Treatment
1 75 YR Female Hospitalization| R ADDR01 IPG, 5076-45 LEAD| ADDR01 IPG, 5076-45 LEAD