REVEAL LINQ
Report
- Report Number
- 9614453-2016-07227
- Event Type
- Malfunction
- Date Received
- December 9, 2016
- Date of Event
- September 21, 2016
- Report Date
- September 21, 2016
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- DSI
- PMA / PMN Number
- K132649
- Removal / Correction Number
- Z-1249-2016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY:THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY SHOWED THE BATTERY INDICATOR SIGNIFYING THAT IT IS TIME FOR DEVICE REPLACEMENT,RECOMMENDED REPLACEMENT TIME (RRT) TRIGGERED ON (B)(6) 2016 DUE TO LOW BATTERY VOLTAGE AFTER BEING IMPLANTED FOR 29 MONTHS. THIS DEVICE WAS REPORTED AS INCLUDED IN THE FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT CANNOT BE CONFIRMED THAT THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. IT IS INCLUDED IN THE FIELD ACTION IN THE ABUNDANCE OF CAUTION.
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) DEVICE REACHED RECOMMENDED REPLACEMENT TIME (RRT) EARLIER THAN EXPECTED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 812053 | REVEAL LINQ | DETECTOR AND ALARM, ARRHYTHMIA | DSI | IPG MFG SWITZERLAND | LNQ11 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |