FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 6161766 · Received December 9, 2016

Report

Report Number
2649622-2016-14790
Event Type
Malfunction
Date Received
December 9, 2016
Date of Event
July 26, 2016
Report Date
September 28, 2016
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. CONCOMITANT PRODUCT: M7700-29 MECHANICAL HEART VALVE IMPLANTED: (B)(6) 1990.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS WERE RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED FOR UNKNOWN REASONS. THE LEADS SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
809110 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407658

Patients

Seq Age Sex Outcome Treatment
1 74 YR ADDR01 IPG, 407652 LEAD