CAPSUREFIX NOVUS
MDR report key: 6161766
·
Received December 9, 2016
Report
- Report Number
- 2649622-2016-14790
- Event Type
- Malfunction
- Date Received
- December 9, 2016
- Date of Event
- July 26, 2016
- Report Date
- September 28, 2016
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. CONCOMITANT PRODUCT: M7700-29 MECHANICAL HEART VALVE IMPLANTED: (B)(6) 1990.
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS WERE RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED FOR UNKNOWN REASONS. THE LEADS SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 809110 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 407658 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | ADDR01 IPG, 407652 LEAD |