5MM PEEK MULTIFUNCTION HANDLE
Report
- Report Number
- 0002936485-2016-01157
- Event Type
- Malfunction
- Date Received
- December 9, 2016
- Date of Event
- November 14, 2016
- Report Date
- April 24, 2017
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- OCZ
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE MANUFACTURE DATE IS NOT KNOWN. ALLEGED FAILURE: I(B)(4) SHAFT INSULATION CRACKED, COMPROMISED, BROKE, OR BREACHED (FAILED INSULSCAN) THE FAILURE(S) IDENTIFIED IN THE INVESTIGATION IS CONSISTENT WITH THE COMPLAINT RECORD. THE PROBABLE ROOT CAUSE/S COULD BE NORMAL WEAR, USER MISUSE, AND IMPROPER STERILIZATION METHODS. THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED. THE DEVICE MANUFACTURE DATE IS NOT KNOWN AT THIS TIME. HOWEVER, SHOULD IT BECOME AVAILABLE IT WILL BE PROVIDED IN FUTURE REPORTS. (B)(4).
IT WAS REPORTED THAT THE INSULATION HAD BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAD BEEN COMPROMISED.