OT VERIO2 METER
Report
- Report Number
- 3008382007-2016-63219
- Event Type
- Malfunction
- Date Received
- December 9, 2016
- Report Date
- December 1, 2016
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K131363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PATIENT
Narratives
THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE REPORTED ISSUE WAS CONFIRMED. ADDITIONAL TESTS WERE PERFORMED ON THE TEST STRIP VIAL AND THE DESICCANT; THESE TESTS PASSED AND FAILED RESPECTIVELY. A SECONDARY ISSUE WAS ALSO OBSERVED DURING TESTING; THE LID OF THE SUBJECT TEST STRIP VIAL HAD A BROKEN BEAK. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING THAT THE CONTROL SOLUTION TEST RESULT OF 167MG/DL WAS ABOVE THE EXPECTED CONTROL SOLUTION RANGE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 811569 | OT VERIO2 METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3947350 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |