FDA Adverse Event Malfunction Summary report: N

OT VERIO2 METER

MDR report key: 6157272 · Received December 9, 2016

Report

Report Number
3008382007-2016-63219
Event Type
Malfunction
Date Received
December 9, 2016
Report Date
December 1, 2016
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K131363
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE REPORTED ISSUE WAS CONFIRMED. ADDITIONAL TESTS WERE PERFORMED ON THE TEST STRIP VIAL AND THE DESICCANT; THESE TESTS PASSED AND FAILED RESPECTIVELY. A SECONDARY ISSUE WAS ALSO OBSERVED DURING TESTING; THE LID OF THE SUBJECT TEST STRIP VIAL HAD A BROKEN BEAK. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING THAT THE CONTROL SOLUTION TEST RESULT OF 167MG/DL WAS ABOVE THE EXPECTED CONTROL SOLUTION RANGE. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
811569 OT VERIO2 METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3947350

Patients

Seq Age Sex Outcome Treatment
1 63 YR