PRODIGY AUTOCODE
Report
- Report Number
- 3005862821-2016-00107
- Event Type
- Injury
- Date Received
- December 9, 2016
- Date of Event
- October 27, 2016
- Report Date
- October 27, 2016
- Manufacturer
- OK BIOTECH CO., LTD.
- Product Code
- NBW
- PMA / PMN Number
- K073118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
SUSPECTED DEVICE EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS. THE STANDBY CURRENT TEST IS 1.0UA. THE CRITERIA IS <55UA. PASS; METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK; TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND RETAIN STRIPS (STRIP LOT NUMBER:D160113-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW ARE 55/56 MG/DL, FOR LEVEL HIGH ARE 243/246 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS; TESTED THE RETURNED STRIPS FROM PATIENT (STRIP LOT NUMBER:D160113-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 59/58 MG/DL; FOR LEVEL HIGH WERE 241/262 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25~70 MG/DL; LEVEL HIGH 210~310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS.
IT WAS REPORTED THAT THE END USER SOUGHT MEDICAL ATTENTION ON (B)(6) 2016 AT 10 PM AFTER RECEIVING A HIGH READING OF 371 MG/DL FROM THE PRODIGY DIABETES GLUCOSE METER. THE END USER FELT WEAK AND THE PARAMEDICS WERE CALLED. THE PARAMEDICS PERFORMED A TEST WITH THEIR METER WITH A RESULT OF 250 MG/DL AND NO TREATMENT WAS PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 808785 | PRODIGY AUTOCODE | BLOOD GLUCOSE MONITORING DEVICE | NBW | OK BIOTECH CO., LTD. | 51850 | 52800-D160113-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Required Intervention | GLIPIZIDE 2.5MG| METFORMIN 500MG| NITRO STAT 0.4 MG |