FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 6156942 · Received December 8, 2016

Report

Report Number
3004753838-2016-57968
Event Type
Malfunction
Date Received
December 8, 2016
Date of Event
November 18, 2016
Report Date
November 18, 2016
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000262
PMA / PMN Number
P120005
Removal / Correction Number
3004753838-02/29/16-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(6) 2016 THAT ON (B)(6) 2016 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. DATA WAS RECEIVED FOR EVALUATION. HOWEVER, INTERMITTENT AUDIO OUTPUT CANNOT BE CONFIRMED THROUGH DATA REVIEW. THE RECEIVER HAS BEEN RECEIVED FOR EVALUATION.  A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.

Description of Event or Problem · 1

THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. A "TRY IT" MANUAL TEST WAS PERFORMED AND SPEAKER DID NOT SOUND. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
806707 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR MDS DEXCOM, INC. MT22719-PNK 5219227 00386270000262

Patients

Seq Age Sex Outcome Treatment
1 43 YR