SYNCHRON LX I 725 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2005-00015
- Event Type
- Other
- Date Received
- June 17, 2005
- Date of Event
- May 19, 2005
- Report Date
- June 17, 2005
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NO INFORMATION
Narratives
A CUSTOMER CONTACTED BECKMAN COULTER REGARDING FALSELY LOW POTASSIUM (K) RESULTS FROM A SINGLE PT THAT WERE GENERATED BY THE SYNCHRON LX I 725 INSTRUMENT. THE CUSTOMER INDICATED THAT AN ER PT SAMPLE WAS TESTED FOR K; THE RESULT WAS 1.7MMOL/L. THE RESULT WAS REPORTED OUT OF THE LAB. THE CUSTOMER INDICATED THAT THE PT WAS ADMINISTERED K BY A BOLUS. THE LAB TESTED THE PT ORIGINAL SAMPLE FOR K ON A DIFFERENT INSTRUMENT IN THEIR LAB. THE K RESULT FROM THE DIFFERENT INSTRUMENT WAS 4.4MMOL/L WAS REPORTED TO THE ER. ON THE NEXT DAY, CUSTOMER COLLECTED A FRESH SAMPLE FROM THE PT AND TESTED IT FOR K ON BOTH INSTRUMENTS (SYNCHRON LX I 725 AND THE DIFFERENT INSTRUMENT). THE SYNCHRON LX I 725 RESULT WAS 2.5MMOL/L, AND THE DIFFERENT INSTRUMENT RESULT WAS 4.6MMOL/L. A 3RD SAMPLE WAS COLLECTED FROM THE PT AND TESTED FOR K ON BOTH INSTRUMENTS. THE SYNCHRON LX I 725 RESULT WAS 2.6MMOL/L. THE DIFFERENT INSTRUMENT RESULT WAS 4.7MMOL/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON LX I 725 CLINICAL SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |