FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 500P AND PAD PAK

MDR report key: 6154220 · Received December 8, 2016

Report

Report Number
3004123209-2016-01819
Event Type
Malfunction
Date Received
December 8, 2016
Date of Event
November 24, 2016
Report Date
July 25, 2017
Manufacturer
HEARTSINE TECHNOLOGIES LTD
Product Code
MKJ
PMA / PMN Number
K041067
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : NOT YET RETURNED TO MANUFACTURER

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). THE DEVICE HISTORY RECORDS FOR THE RETURNED SAM 300P DEVICE WAS REVIEWED AND THIS CONFIRMED THAT ALL MANUFACTURING AND QUALITY CHECKS AND TESTS HAD BEEN SUCCESSFULLY COMPLETED PRIOR TO THE DESPATCH OF THE SAM 300P FROM HEARTSINE TECHNOLOGIES, BELFAST ON THE (B)(6) 2014. THE HISTORY LOG FOR THIS DEVICE SHOWED THAT THE PAD-PAK WAS FIRST INSTALLED ON THE (B)(6) 2014 AND THAT IT HAD PERFORMED TO SPECIFICATION UP TO THE AUTO SELF-TEST ON THE (B)(6) 2016. THE DEVICE WAS DISASSEMBLED FOR INVESTIGATION AND THE INVESTIGATION FOUND THAT AN INTERNAL COMPONENT HAD FAILED. THE FAILURE OF THIS COMPONENT MOST LIKELY LED TO AN EXCESS CURRENT DRAIN, RESULTING IN THE DEPLETION OF THE COIN CELL. THE DEPLETION OF THIS COIN CELL HAD RESULTED IN THE PAD CLOCK FAILING TO INCREMENT UPON RECEIPT AT HEARTSINE. THE USER WOULD HAVE BEEN ALERTED TO THIS WITH A "WARNING, DEVICE SERVICE REQUIRED" PROMPT AND BY A CONSTANTLY LIT STATUS LED.

Description of Event or Problem · 0

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE SERVICE REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
807756 HEARTSINE SAMARITAN 500P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES LTD SAM N/A

Patients

Seq Age Sex Outcome Treatment
1