FDA Adverse Event
Malfunction
Summary report: N
S8-3T MICRO TEE TRANSDUCER
MDR report key: 6152758
·
Received December 7, 2016
Report
- Report Number
- 3019216-2016-00045
- Event Type
- Malfunction
- Date Received
- December 7, 2016
- Report Date
- November 22, 2016
- Manufacturer
- PHILIPS ULTRASOUND, INC
- Product Code
- ITX
- PMA / PMN Number
- K030455
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
EVALUATION OF THE TRANSDUCER WILL BE INCLUDED IN A FOLLOW UP REPORT UPON ITS RETURN AND INVESTIGATION COMPLETION.
Additional Manufacturer Narrative · 1
EVALUATION OF THE TRANSDUCER CONFIRMED OIL SEPARATION IN THE WINDOW, DAMAGE TO THE WINDOW, AND A DENTED SHAFT. THE CAUSE OF THIS DAMAGE IS INDICATIVE OF IMPROPER HANDLING AND MAINTENANCE.
Description of Event or Problem · 1
A CUSTOMER REPORTED AN S8-3T MODEL TRANSDUCER WAS PRODUCING A WEAK IMAGE. THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT.