ATTUNE PATELLA DRILL CLAMP
Report
- Report Number
- 1818910-2016-32985
- Event Type
- Malfunction
- Date Received
- December 7, 2016
- Date of Event
- December 5, 2016
- Report Date
- December 5, 2016
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- HXD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
ADDITIONAL NARRATIVE: THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. FOLLOW UP COMMUNICATION FOR PRODUCT RETURN WAS UNSUCCESSFUL. THE PRODUCT DEVELOPMENT TEAM IS AWARE OF THIS COMPLAINT AND HAS A NEW DESIGN TO MITIGATE THE REPORTED FAILURE. PREVIOUS INVESTIGATIONS AND EVALUATIONS FOUND THAT THE FAILURE MAY HAVE OCCURRED WHILE THE CLAMP WAS BEING SQUEEZED UNDER VERY HIGH LOADS AND THE BEARINGS INSIDE OF THE CYLINDER WERE EXTRUDING SLIGHTLY, CAUSING THE MECHANISM TO STICK. THE ENHANCED ATTUNE PATELLA DRILL CLAMP WILL BE DISTRIBUTED UNDER PRODUCT CODE 254501055. (B)(4) WAS APPROVED ON NOVEMBER 25, 2014. NEW DESIGN IS NOW AVAILABLE. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THE CLAMP WAS GETTING STUCK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 804240 | ATTUNE PATELLA DRILL CLAMP | KNEE INSTRUMENT/TRIAL | HXD | DEPUY ORTHOPAEDICS, INC. | NW1462664 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |