FDA Adverse Event Malfunction Summary report: N

SYNCARDIA COMPANION 2 DRIVER

MDR report key: 6151496 · Received December 7, 2016

Report

Report Number
3003761017-2016-00384
Event Type
Malfunction
Date Received
December 7, 2016
Date of Event
November 28, 2016
Report Date
November 28, 2016
Manufacturer
SYNCARDIA SYSTEMS, LLC
Product Code
LOZ
PMA / PMN Number
P030011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPANION 2 DRIVER WAS RETURNED TO SYNCARDIA FOR EVALUATION. REVIEW OF THE PATIENT FILE REVEALED MULTIPLE SYSTEM CHECK FAILURES, WHICH ALIGNS WITH THE STATEMENTS MADE BY THE CUSTOMER. DURING INVESTIGATION TESTING, THE DRIVER FUNCTIONED AS INTENDED AND THERE WAS NO EVIDENCE OF A MALFUNCTION. THE CUSTOMER-REPORTED SYSTEM CHECK FAILURES WERE REPRODUCED BY OBSTRUCTING THE DRIVELINE PORT WITH ORANGE SHIPPING CAPS. BASED ON PREVIOUS EXPERIENCE, THIS USE ERROR IS THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT. THIS ISSUE WILL CONTINUE TO BE MONITORED AND TRENDED AS PART OF THE CUSTOMER EXPERIENCE PROCESS. SYNCARDIA HAS COMPLETED ITS EVALUATION OF THIS COMPLAINT AND IS CLOSING THIS FILE. CE 3602 FOLLOW-UP REPORT 1.

Additional Manufacturer Narrative · 1

THIS ALLEGED FAILURE MODE POSES A LOW RISK TO A PATIENT BECAUSE THE ISSUE WAS OBSERVED WHEN THE COMPANION 2 DRIVER WAS NOT SUPPORTING A PATIENT. THE COMPANION 2 DRIVER HAS BEEN RETURNED TO SYNCARDIA FOR EVALUATION. THE RESULTS OF THE EVALUATION WILL BE PROVIDED IN A FOLLOW-UP MDR. (B)(4).

Description of Event or Problem · 1

THE COMPANION 2 DRIVER WAS NOT SUPPORTING A PATIENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE COMPANION 2 DRIVER DID NOT PASS THE SYSTEM CHECKOUT.

Description of Event or Problem · 1

THE COMPANION 2 DRIVER WAS NOT SUPPORTING A PATIENT. THE CUSTOMER, A SYNCARDIA CERTIFIED HOSPITAL, REPORTED THAT THE COMPANION 2 DRIVER DID NOT PASS THE SYSTEM CHECKOUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
801371 SYNCARDIA COMPANION 2 DRIVER EXTERNAL PNEUMATIC DRIVER LOZ SYNCARDIA SYSTEMS, LLC

Patients

Seq Age Sex Outcome Treatment
1