FDA Adverse Event Malfunction Summary report: N

MAQUET MODULAR UNIVERSAL TABLE TOP

MDR report key: 6150823 · Received December 7, 2016

Report

Report Number
8010652-2016-00030
Event Type
Malfunction
Date Received
December 7, 2016
Date of Event
October 30, 2016
Report Date
February 28, 2017
Manufacturer
KEHLER STRASSE 31
Product Code
FQO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TABLE TOP HAS BEEN SEND TO (B)(4) FOR INVESTIGATION AND REPAIR. DURING THE INVESTIGATION MAQUET FOUND THAT THE CENTRAL HOUSING OF THE TABLE TOP WAS BROKEN ON TWO LOCATIONS TOWARD THE HEAD SIDE OF THE TABLE TOP. COLLISION WITH AN OBSTACLE BELOW THE TABLE TOP WHEN ADJUSTING THE HEIGHT (FOR EXAMPLE WHEN THE TRANSPORTER IS NOT MOVED COMPLETELY UNDER THE COLUMN FOR THE POSTOPERATIVE TRANSFER) CAN CAUSE THIS TYPE OF REPORTED MALFUNCTION. MAQUET FOUND MARKS ON THE UNDERSIDE OF THE TABLE TOP CONSISTENT WITH COLLISION DAMAGE. MAQUET (B)(4) PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Additional Manufacturer Narrative · 1

THE TABLE TOP IN QUESTION WAS REQUESTED FOR INVESTIGATION. AT THE TIME OF THIS REPORT HOWEVER THE TABLE TOP HAS NOT YET BEEN ARRIVED SO THAT THE INVESTIGATION HAS NOT YET BEEN CARRIED OUT. MAQUET (B)(4) PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS, AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Description of Event or Problem · 1

(B)(4).

Description of Event or Problem · 1

A PATIENT WAS POSITIONED ON THE TABLE TOP. DURING TRANSFER OF THE TABLE TOP TO THE COLUMN, THE TABLE TOP WAS MOVED TO THE HEAD END. THIS RESULTED IN AN OVERLOAD AND BREAK OF THE TABLE TOP AT THE MOUNTING. THE TABLE TOP TILTED FORWARD, AND WAS INTERCEPTED BY THE STAFF SO THAT NO PATIENT INJURY OCCURRED. NO INJURY WAS REPORTED TO MAQUET. (B)(4).