PERMOBIL CHAIRMAN ENTRA
Report
- Report Number
- 1221084-2016-00051
- Event Type
- Death
- Date Received
- December 6, 2016
- Report Date
- January 4, 2017
- Manufacturer
- PERMOBIL AB (PAB)
- Product Code
- ITI
- PMA / PMN Number
- K960951
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
UNIT WAS INSPECTED JOINTLY BY PERMOBIL (B)(4), AND NO ISSUES WERE NOTED THAT INDICATE THE CHAIR TO HAVING MALFUNCTIONED WHICH WOULD HAVE LED TO THE EVENT. REPORTS RECEIVED WERE THE UNIT WAS FOUND FULLY OPERATIONAL WITH NO RECORDED FAULTS IN THE SYSTEM LOG INDICATING A SYSTEM MALFUNCTION THAT COULD HAVE ATTRIBUTED TO THE REPORTED EVENT. THE ACTUAL ROOT CAUSE OF THIS EVENT CANNOT BE DETERMINED, BUT IT IS SPECULATED THAT IT WAS EITHER USER OR ATTENDANT DRIVE ERRORS THAT LED TO THE REPORTED EVENT.
PERMOBIL RECEIVED REPORT FROM (B)(6) POLICE - (B)(6) THAT THEY WERE INVESTIGATING AN INCIDENT RESULTING IN AN INDIVIDUAL'S DEATH. REPORTS GIVEN WERE AN INDIVIDUAL WAS BEING LOADED INTO A VEHICLE WHILE OCCUPYING A PERMOBIL WHEELCHAIR. AS CLIENT WAS BEING MANEUVERED FROM THE LIFT PLATFORM INTO THE VEHICLE, THE WHEELCHAIR ALLEGEDLY DROVE IN THE OPPOSITE DIRECTION AS INTENDED, OFF OF THE END OF THE PLATFORM LIFT, RESULTING IN THE END-USERS DEATH. IT WAS NOT REPORTED WHEN THE ACTUAL DATE THIS EVENT OCCURRED OR WHEN THEREAFTER THE END-USER DECEASED. IT IS UNCLEAR AT THIS TIME IF THE CHAIR MALFUNCTIONED, OR IF THIS EVENT IS ATTRIBUTED TO USER/ATTENDANT ERROR. AN INSPECTION OF THE CHAIR HAS NOT BEEN PERFORMED AT THIS TIME DUE TO THE ONGOING POLICE INVESTIGATION. AN INSPECTION DATE HAS BEEN SCHEDULED FOR A (B)(4) INVESTIGATION FIRM, ALONG WITH PERMOBIL, TO INSPECT THE CHAIR FOR ANY IRREGULARITIES. UPON THE COMPLETION OF THE INSPECTION AND INVESTIGATION, A FOLLOW-UP MDR WILL BE SUBMITTED.
RECEIVED REPORT THAT DRIVER ATTENDANT WAS PRELOADING THE WHEELCHAIR ON A VEHICLE LIFT WHILE THE END-USER WAS OCCUPYING THE WHEELCHAIR. IN AN ATTEMPT TO MANEUVER THE WHEELCHAIR INTO THE VEHICLE, THE DEVICE ALLEGEDLY DROVE BACKWARDS OVER THE EDGE OF THE VEHICLE LIFT. REPORTS ARE CLIENT SUBSEQUENTLY DIED AS A RESULT OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 799941 | PERMOBIL CHAIRMAN ENTRA | POWERED WHEELCHAIR | ITI | PERMOBIL AB (PAB) | CHAIRMAN ENTRA | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |