FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 6149274 · Received December 6, 2016

Report

Report Number
1000113657-2016-01960
Event Type
Malfunction
Date Received
December 6, 2016
Date of Event
November 10, 2016
Report Date
December 19, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292007041
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECT FOUND ON RETURNED METER. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER'S TEST STRIPS HAD A POOR STORAGE(KITCHEN). (B)(4).

Additional Manufacturer Narrative · 1

(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) (B)(4). PRODUCT DOES NOT RETURN FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER'S TEST STRIPS HAD A POOR STORAGE(KITCHEN). NOTE: TEST STRIP-(B)(4).

Description of Event or Problem · 1

COSTUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 181, 177 AND 186 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100-130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 195 AND 209 MG/DL. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION,CUSTOMER STORES THE PRODUCT IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/16/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(4). MEMORY CONCERNS: ALL RESULTS FROM METER MEMORY. CUSTOMER HAS NOT HAD ANY RECENT CHANGES IN HER DIET, EXERCISE, OR MEDICATIONS. CUSTOMER INFORMED OF PROPER SUPPLY STORAGE.

Description of Event or Problem · 1

COSTUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 181, 177 AND 186 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100-130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 195 AND 209 MG/DL. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION,CUSTOMER STORES THE PRODUCT IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 12/16/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6). MEMORY CONCERNS: ALL RESULTS FROM METER MEMORY. CUSTOMER HAS NOT HAD ANY RECENT CHANGES IN HER DIET, EXERCISE, OR MEDICATIONS. CUSTOMER INFORMED OF PROPER SUPPLY STORAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
800606 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1850 00021292007041

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY