FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 6146534 · Received December 5, 2016

Report

Report Number
2954323-2016-07483
Event Type
Malfunction
Date Received
December 5, 2016
Date of Event
November 14, 2016
Report Date
April 4, 2017
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
NBW
PMA / PMN Number
K092602
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. (B)(6).

Additional Manufacturer Narrative · 1

THIS SERVES AS A CORRECTION REPORT. BRAND NAME WAS INCORRECTLY IDENTIFIED AS FREESTYLE FREEDOM LITE. THE CORRECT BRAND NAME FOR THE REPORTED METER SERIAL NUMBER IS FREESTYLE LITE. IN ADDITION PMA/510K# HAS BEEN UPDATED TO K092602, AND CATALOG NO# HAS BEEN UPDATED TO 71501-01.

Additional Manufacturer Narrative · 1

THE RETURNED METER WAS INVESTIGATED WITH RETAINED TEST STRIPS. METER POWERED ON WITH BUTTON DEPRESSION AND TEST STRIP INSERTION. AN ERROR MESSAGE WAS OBSERVED UPON BUTTON DEPRESSION AND TEST STRIP INSERTION. THE REPORTED COMPLAINT WAS NOT CONFIRMED.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
796891 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW ABBOTT DIABETES CARE INC 71501-01

Patients

Seq Age Sex Outcome Treatment
1