SCREWDRIVER / DEPTH GAUGE 2 IN 1 AUTOFIX 2.0/2.5, 10-30MM
Report
- Report Number
- 0008031020-2016-00587
- Event Type
- Malfunction
- Date Received
- December 5, 2016
- Date of Event
- November 17, 2016
- Report Date
- April 26, 2017
- Manufacturer
- STRYKER GMBH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
THE REPORTED EVENT THAT THE TIP OF A SCREWDRIVER / DEPTH GAUGE 2 IN 1 AUTOFIX 2.0/2.5, 10-30MM BROKE DURING SURGERY COULD BE CONFIRMED. BASED ON INVESTIGATION, THE ROOT CAUSE WAS ATTRIBUTED TO BE USER RELATED. THE MOST LIKELY CAUSE IS THAT EXCESSIVE TORQUE WAS APPLIED DURING SCREWING. IN ADDITION TO THIS, ALSO THE FOLLOWING FACTORS MIGHT HAVE CONTRIBUTED TO THE EVENT: - THE INSTRUMENT WAS NOT EXAMINED FOR WEAR AND TEAR BEFORE SURGERY. THE DEVICE HAD EXPERIENCED EXCESSIVE USE OR FORCE AND BECAME SUSCEPTIBLE TO BREAKAGE. - HARD/DENSE BONE. THE DEVICE INSPECTION REVEALED THE FOLLOWING: THE DEVICE WAS RECEIVED WITH BROKEN TIP. NO FRAGMENTS HAVE BEEN PROVIDED. THE HELICAL TREND OF THE BREAKAGE IS TYPICAL OF OVERTORQUE. THE FRACTURE BREAKAGE WAS ANALYZED. MOST OF IT PRESENTS THE PATTERN OF INSTANTANEOUS ZONE (UNIFORM FINE GRAINED TEXTURE), REPRESENTING THE FINAL FRACTURE ZONE (OVERLOAD FRACTURE). A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT WAS PERFORMED. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO INDICATIONS OF MATERIAL RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION. IF ANY FURTHER INFORMATION IS PROVIDED, THE INVESTIGATION REPORT WILL BE UPDATED.
DURING THE TREATMENT FOR A BROKEN SCAPHOID. INSERTED THE GW INTO BROKEN BONE AREA AND MEASURED A SCREW LENGTH BY THE DEPTH GAUGE WITH DRIVER. THE TIP OF THE DRIVER BROKE DURING INSERTION OF THE SCREW. THE BROKEN PART WAS REMOVED UNDER X RAY. REMOVED THE GW AND INSERTED THE SCREW BY DRIVER SHAFT FOR REMOVING, AND COMPLETE THE PROCEDURE.
DURING THE TREATMENT FOR A BROKEN SCAPHOID. INSERTED THE GW INTO BROKEN BONE AREA AND MEASURED A SCREW LENGTH BY THE DEPTH GAUGE WITH DRIVER. THE TIP OF THE DRIVER BROKE DURING INSERTION OF THE SCREW. THE BROKEN PART WAS REMOVED UNDER X RAY. REMOVED THE GW AND INSERTED THE SCREW BY DRIVER SHAFT FOR REMOVING, AND COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 794897 | SCREWDRIVER / DEPTH GAUGE 2 IN 1 AUTOFIX 2.0/2.5, 10-30MM | INSTRUMENT | LXH | STRYKER GMBH | 77178 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |