FDA Adverse Event Injury Summary report: N

DOUBLE MOBILITY HC LINER 28/DMD

MDR report key: 6141272 · Received December 2, 2016

Report

Report Number
3005180920-2016-00631
Event Type
Injury
Date Received
December 2, 2016
Date of Event
November 3, 2016
Report Date
January 4, 2017
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
MEH
PMA / PMN Number
K092265
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 12 DECEMBER 2016 THE R&D PROJECT MANAGER PERFORMED A VISUAL INSPECTION OF THE RETRIEVED EXPLANTS AND COMMENTED AS FOLLOWS: THE PIECES WERE ANALYZED. THE CERAMIC HEAD WAS INSIDE THE DM LINER. SOME SIGNS INSIDE THE HEAD WERE NOTICED, PROBABLY DUE TO THE GRIP WITH THE TAPER. A LITTLE SIGN WAS PRESENT ON THE RIM OF THE HEAD, PROBABLY OCCURRED DURING THE EXTRACTION. FROM THE RECEIVED PIECES IT WAS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE EVENT.

Additional Manufacturer Narrative · 1

BATCH REVIEWS PERFORMED ON (B)(6) 2016. LOT 160828: (B)(4) ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2016. EXPIRATION DATE: 2021-05-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE L +3.5, CODE 01.29.203, LOT. 161742 ((B)(4)). A (B)(4) ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2016. EXPIRATION DATE: 2021-04-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT YET RECEIVED.

Description of Event or Problem · 1

THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE SURGEON PERFORMED AN I & D AND SWAPPED THE HEAD AND THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
793233 DOUBLE MOBILITY HC LINER 28/DMD DOUBLE MOBILITY LINER MEH MEDACTA INTERNATIONAL SA 161392

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention