COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS
Report
- Report Number
- 1820334-2016-01352
- Event Type
- Malfunction
- Date Received
- December 1, 2016
- Date of Event
- October 30, 2016
- Report Date
- August 17, 2017
- Manufacturer
- COOK INC
- Product Code
- OQY
- UDI-DI
- 10827002242378
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE EVENT IS CURRENTLY UNDER INVESTIGATION.
INVESTIGATION - EVALUATION: THE COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS WAS NOT RETURNED/RECEIVED FOR AN EVALUATION. AS REPORTED BY THE CUSTOMER, THE DEVICE WAS DISCARDED AT THE USER FACILITY. THE REPORTED BALLOON LEAK COULD ONLY BE CONFIRMED BASED ON THE CUSTOMER¿S TESTIMONY. BASED ON THE INFORMATION AVAILABLE A DEFINITIVE ROOT CAUSE COULD NOT BE ESTABLISHED. A REVIEW OF PRODUCTION AND QUALITY DOCUMENTATION DID NOT IDENTIFY ANY SPECIFIC ISSUES WITH CURRENT MANUFACTURING OR QUALITY CONTROLS THAT MAY HAVE CONTRIBUTED TO THIS INCIDENT. A REVIEW OF THE DEVICE HISTORY RECORD SHOWED THERE WERE NO NON-CONFORMANCES IDENTIFIED DURING THE MANUFACTURING PROCESS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED PRODUCT ISSUE. A REVIEW OF COMPLAINT HISTORY FOR THIS PRODUCT AND LOT NUMBER COMBINATION REVEALED THIS IS THE ONLY COMPLAINT ASSOCIATED WITH THIS LOT NUMBER. PER THE QUALITY ENGINEERING RISK ASSESSMENT; NO FURTHER ACTION IS REQUIRED. COOK MEDICAL HAS NOTIFIED THE APPROPRIATE PERSONNEL AND THE WILL CONTINUE TO MONITOR THIS DEVICE VIA THE COMPLAINTS DATABASE FOR SIMILAR COMPLAINTS.
IT WAS REPORTED THAT THE OPERATOR PLACED THE BALLOON IN THE PATIENT, WHEN THEY BEGAN INJECTING THE SALINE, FOUND THE BALLOON WAS BROKEN AND LEAKING, THEN THE BALLOON WAS REMOVED AND ANOTHER NEW ONE WAS USED TO FINISH THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 787991 | COOK BAKRI POSTPARTUM BALLOON WITH RAPID INSTILLATION COMPONENTS | OQY INTRAUTERINE BALLOON | OQY | COOK INC | N/A | 10827002242378 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |