FDA Adverse Event Injury Summary report: N

EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE

MDR report key: 6137125 · Received November 30, 2016

Report

Report Number
2951238-2016-00900
Event Type
Injury
Date Received
November 30, 2016
Report Date
March 3, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDS
PMA / PMN Number
PK100584
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE SERIAL NUMBER OF THE GASTROSCOPE INVOLVED INTO THE REPORT, AND TO PROVIDE INSTRUMENTS HISTORY REVIEW FOR THE DEVICE. THE GASTROSCOPE WAS RETURNED TO OLYMPUS FOR SERVICE ON MARCH 08, 2016 FOR A CHANNEL REPLACEMENT AS PER CUSTOMER'S REQUEST. THE EVALUATION FINDINGS REVEALED A CRACKED LIGHT GUIDE LENSES, SCRATCHED INSERTION TUBE, AND DEEP SCRATCHES NOTED ON THE CONTROL BODY. THE DEVICE WAS SERVICED AND RETURNED TO THE USER FACILITY.

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO MAKE A CORRECTION ON THE PROCODE FROM GCK TO FDS.

Additional Manufacturer Narrative · 1

OLYMPUS CANNOT DETERMINE IF THE GASTROSCOPE REFERENCED IN THE ARTICLE HAS BEEN RETURNED TO OLYMPUS FOR EVALUATION OR NOT, AS THERE WAS NO SPECIFIC SERIAL NUMBER OF GASTROSCOPE WAS PROVIDED IN THE ARTICLE. THE CAUSE OF THE REPORTED INCIDENTS CANNOT BE DETERMINED AT THIS TIME, BUT IMPROPER MAINTENANCE OF THE DEVICE CANNOT BE RULED OUT AS A CONTRIBUTORY FACTOR TO THE REPORTED EVENTS. OLYMPUS WILL CONTINUE TO INVESTIGATE THIS REPORT AND WILL UPDATE THE REPORT IF ADDITIONAL INFORMATION BECOMES AVAILABLE AT A LATER TIME.

Description of Event or Problem · 1

ON NOVEMBER 1, 2016 OLYMPUS RECEIVED A CLINICAL ARTICLE TITLED, ¿CARBAPENEM-RESISTANT KLEBSIELLA PNEUMONIAE (CR-KP) CLUSTER ASSOCIATED WITH GASTROSCOPE EXPOSURE AMONG SURGICAL INTENSIVE CARE UNIT (SICU) PATIENTS UPMC. THE ARTICLE SHOWED THAT IN JANUARY 2016 AN INCREASE IN CR-KP ISOLATES WAS OBSERVED IN THE OPEN 8 BED SICU AT THE FACILITY. A TOTAL OF SIX CASES WERE IDENTIFIED IN JANUARY AND ALL SIX CASES WERE EXPOSED TO THE SAME OR GASTROSCOPE. AS A RESULT, THE FACILITY CONDUCTED AN INVESTIGATION TO DETERMINE THE EXPOSURE SOURCE AS WELL AS THE RELATEDNESS OF CR-KP STRAINS. THE USER FACILITY COLLECTED RECTAL SWABS ON ALL CURRENT IN-PATIENTS EXPOSED TO THE SICU IN JANUARY 2016 TO ASSESS FOR CRE CARRIAGE. ASIDE FROM THE SIX CASES, NO PATIENTS EXPOSED TO THE GASTROSCOPES WERE STILL IN PATIENT AT THE TIME OF THE INVESTIGATION. THE GASTROSCOPE WAS REMOVED FROM SERVICE AND REPROCESSING RECORDS WERE REVIEWED. THE FACILITY CONDUCTED A MICROBIOLOGIC ASSESSMENT UTILIZING TWO SCOPE CULTURE TECHNIQUES. A BOROSCOPY WAS PERFORMED TO EXAMINE THE SCOPE LUMEN FOR DEFECTS. THE FACILITY STAFF MOVED PATIENTS OUT OF THE SICU TO THOROUGHLY CLEAN SPACE AND CONDUCT ULTRAVIOLET LIGHT DISINFECTION. THE CR-KP ISOLATES WERE SENT TO MICROBIAL GENOME EPIDEMIOLOGY LABORATORY (MIGEL) FOR PULSED-FIELD GEL ELECTROPHORESIS (PFGE) TYPING AND WHOLE GENOME SEQUENCING (WGS) TO ASSESS RELATEDNESS. APPROX 43 ADDITIONAL PATIENTS WERE REPORTEDLY EXPOSED TO SICU IN JANUARY, AND 19 OUT OF THE 43 WERE STILL IN HOUSE AND AVAILABLE FOR RECTAL SCREENING. ONLY 17 OUT OF 19 PATIENTS HAD UNDERGONE RECTAL SWABBING, AS TWO PATIENTS WERE DISCHARGED PRIOR TO OBTAIN THE SWAB. ONE OF THE PATIENT¿S RECTAL SWAB TESTED POSITIVE FOR CR-KP BUT THIS PATIENT WAS NEVER EXPOSED TO THE GASTROSCOPE. AS A RESULT OF THE POSITIVE RECTAL SWAB, THE FACILITY CONTINUED THE WEEKLY SCREENING UNTIL NO NEW POSITIVE RESULTS, AND SUBSEQUENT TESTING WERE NEGATIVE FOR CRE; THEREFORE, THE WEEKLY SCREENING WAS STOPPED. THE ARTICLE SHOWED THAT BASED ON THE ASSESSMENT CONDUCTED ON THE GASTROSCOPE THE FOLLOWING INFORMATION WERE REPORTED: NO DEFICIENCIES IDENTIFIED WITH SCOPE REPROCESSING, NO ORGANISMS IDENTIFIED USING EITHER OF THE SCOPE CULTURING METHODS, BOROSCOPY OF THE SCOPE LUMEN REVEALED DEEP SCRATCHES AND LUMINAL DEBRIS, AFTER BOROSCOPY THE SCOPE LUMEN WAS REMOVED AND REPLACED WITH NEW LUMEN BEFORE RETURNING TO SERVICE. THE GASTROSCOPE WAS REPORTEDLY USED ON A PATIENT WITH A KNOWN CR-KP AT THE END OF (B)(6) 2015. THE ARTICLE SHOWED THAT A CR-KP TRANSMISSION LIKELY OCCURRED THROUGH A CONTAMINATED GASTROSCOPE. REPORT 1 OF 6 REPORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786348 EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GASTROINTESTINAL VIDEOSCOPE FDS OLYMPUS MEDICAL SYSTEMS CORP. GIF-H180 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other