FDA Adverse Event Malfunction Summary report: N

MCKESSON RADIOLOGY

MDR report key: 6136666 · Received November 30, 2016

Report

Report Number
8022257-2016-00003
Event Type
Malfunction
Date Received
November 30, 2016
Date of Event
November 16, 2016
Report Date
November 30, 2016
Manufacturer
MCKESSON MEDICAL IMAGING COMPANY
Product Code
LLZ
PMA / PMN Number
K140909
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MCKESSON HAS IDENTIFIED A SOFTWARE DEFECT THAT IS ENCOUNTERED WHEN USERS LOGIN TO THE PACS FROM AN EMR, OR USING LEGACY WEB URL. AS A RESULT OF THE SOFTWARE DEFECT, THE USER MAY EXPERIENCE MISSING IMAGES IN A NEWLY IMPORTED STUDY, AND/OR STUDY IMPORTS THAT REMAIN IN AN "IN-PROGRESS" STATUS. MCKESSON WILL WORK WITH AFFECTED CUSTOMERS TO APPLY A SOFTWARE UPDATE TO PREVENT THE RECURRENCE OF THE PROBLEM.

Description of Event or Problem · 1

THE REPORTING FACILITY HAD REPORTED A ONE HOUR DELAY IN CARE FOR A (B)(6) YEAR OLD STROKE PATIENT DUE TO A FAILURE TO IMPORT ALL STUDY IMAGES INTO THE PACS SYSTEM. THE REPORTING FACILITY PERFORMED ALTERNATIVE MEASURES TO VIEW THE STUDY IMAGES TO TREAT THE STROKE PATIENT. NO HARM TO THE PATIENT WAS REPORTED AS A RESULT OF THE DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786582 MCKESSON RADIOLOGY PICTURE ARCHIVING AND COMMUNICATION SYSTEM LLZ MCKESSON MEDICAL IMAGING COMPANY 12.2

Patients

Seq Age Sex Outcome Treatment
1 27 YR