FDA Adverse Event Malfunction Summary report: N

AED PLUS

MDR report key: 6136410 · Received November 30, 2016

Report

Report Number
1220908-2016-02958
Event Type
Malfunction
Date Received
November 30, 2016
Report Date
November 10, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K011541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS ATTRIBUTED TO A FAULTY INTEGRATED CIRCUIT ON THE MAIN BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
784787 AED PLUS DEFIBRILLATOR MKJ ZOLL MEDICAL CORPORATION AED PLUS NA

Patients

Seq Age Sex Outcome Treatment
1