FDA Adverse Event Injury Summary report: N

SIGN IM NAIL

MDR report key: 6136312 · Received November 30, 2016

Report

Report Number
3034525-2016-00302
Event Type
Injury
Date Received
November 30, 2016
Date of Event
December 10, 2016
Report Date
December 16, 2016
Manufacturer
SIGN FRACTURE CARE INTERNATIONAL
Product Code
HSB
PMA / PMN Number
K022632
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BG
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A PRODUCT INVESTIGATION WAS PERFORMED FOR THIS DEVICE. THE ACTUAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. THE ROOT CAUSE OF THE BROKEN NAIL IS UNDETERMINED. THE RADIOGRAPHIC AND CLINICAL DATA WERE REVIEWED BY A SIGN ORTHOPEDIC SURGEON. CORRECTIVE SURGERY HAS NOT BEEN SCHEDULED AS OF THIS DATE. SIGN FRACTURE CARE INTERNATIONAL CONTINUES TO MONITOR THESE EVENTS AS PART OF OUR POST MARKET ACTIVITIES. A FOLLOW UP REPORT WILL BE FILED WHEN WE RECEIVE NEW INFORMATION REGARDING THIS CASE.

Additional Manufacturer Narrative · 1

A PRODUCT INVESTIGATION WAS PERFORMED FOR THIS DEVICE. THE ACTUAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. THE ROOT CAUSE OF THE BROKEN NAIL IS UNDETERMINED. THE RADIOGRAPHIC AND CLINICAL DATA WERE REVIEWED BY A SIGN ORTHOPEDIC SURGEON. THERE IS NO WAY TO PREDICT A NON-UNION OR FAILURE TO HEAL. SIGN FRACTURE CARE INTERNATIONAL CONTINUES TO MONITOR THESE EVENTS AS PART OF OUR POST MARKET ACTIVITIES.

Description of Event or Problem · 1

WE BECAME AWARE ON (B)(6) 2016, IT WAS REPORTED THAT A SIGN IM NAIL IMPLANTED TO REPAIR A FRACTURE WAS FOUND TO BE BROKEN DURING A FOLLOW UP VISIT.

Description of Event or Problem · 1

WE BECAME AWARE ON (B)(6)2016 SIGN IM NAIL IMPLANTED TO REPAIR A FRACTURE WAS EXCHANGED. THE IM NAIL WAS PREVIOUSLY REPORTED TO THE FDA AS BROKEN ON (B)(6) 2016, WITH THE PATIENT IDENTIFIER (B)(6). THE BROKEN IM NAIL WAS REPLACED WITH A 10MM X 320MM STANDARD NAIL PER THE SIGN TECHNIQUE MANUAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786856 SIGN IM NAIL INTRAMEDULLARY FIXATION ROD HSB SIGN FRACTURE CARE INTERNATIONAL

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R