FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO 2 WITH VASOSHIELD

MDR report key: 6135832 · Received November 30, 2016

Report

Report Number
2242352-2016-01217
Event Type
Malfunction
Date Received
November 30, 2016
Date of Event
November 2, 2016
Report Date
November 3, 2016
Manufacturer
MAQUET CV
Product Code
GEI
PMA / PMN Number
K101274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WERE OBSERVED. A VISUAL INSPECTION DETERMINED THAT THE HEATER WIRE WAS SLIGHTLY FLEXED AWAY AT THE CENTER FROM THE HOT JAW. THE WIRE REMAINED IN PLACE AT THE BASE AND TIP OF THE JAWS. TISSUE AND CHAR BUILDUP WAS OBSERVED ON THE JAWS. BASED ON THE RETURNED CONDITION OF THE DEVICE THE COMPLAINT WAS CONFIRMED FOR REPORTED FAILURE MODE ¿BENT WIRE.¿ SPECIFIC ACTIONS FOR THE REPORTED FAILURE MODE ARE BEING MAINTAINED AND DOCUMENTED UNDER MAQUET'S FAILURE INVESTIGATION REPORT (FIR) SYSTEM.

Additional Manufacturer Narrative · 1

(B)(4). A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. THE DEVICE HAS BEEN RETURNED TO THE FACTORY AND IS BEING EVALUATED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIEW HEMOPRO 2 WITH VASOSHIELD METAL PART OF ONE SIDE OF THE JAWS BENT. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, VASOVIEW HEMOPRO 2 WITH VASOSHIELD METAL PART OF ONE SIDE OF THE JAWS BENT. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
787720 VASOVIEW HEMOPRO 2 WITH VASOSHIELD ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI MAQUET CV 25128289

Patients

Seq Age Sex Outcome Treatment
1