O-ARM 1000 IMAGING SYSTEM
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 1723170-2016-04286
- Event Type
- Malfunction
- Date Received
- November 30, 2016
- Date of Event
- December 3, 2015
- Report Date
- November 30, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K050996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. A SITE REPRESENTATIVE REPORTED THAT WHEN BOOTING THE IMAGING SYSTEM THE MOBILE VIEW STATION (MVS) WAS NOT CONNECTING. NO PATIENT WAS PRESENT DURING THE TIME OF THE REPORTED INCIDENT. AN ONSITE INSPECTION WAS SCHEDULED FOR A LATER DATE. DURING THE ONSITE INSPECTION, THE MEDTRONIC REPRESENTATIVE REPORTED THAT THE FOLLOWING PARTS WERE REPLACED: THE IMAGING SYSTEM COMPUTER, THE PCBA SWITCHING BOARD, AND THE MVS COMPUTER. THE REPLACED PARTS WERE SENT TO THE MANUFACTURER FOR PART ANALYSIS. A SUCCESSFUL SYSTEM CHECKOUT WAS PERFORMED WHICH VERIFIED THAT THE ISSUE HAD BEEN RESOLVED. DURING PART ANALYSIS THERE WAS A CONFIRMED SOFTWARE FAILURE WITH THE IMAGING SYSTEM COMPUTER. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA (B)(4) OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 FDA-483 FEI: (B)(4) . THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
A SITE REPRESENTATIVE REPORTED THAT WHEN BOOTING THE IMAGING SYSTEM THE MOBILE VIEW STATION (MVS) WAS NOT CONNECTING. NO PATIENT WAS PRESENT DURING THE TIME OF THE REPORTED INCIDENT. AN ONSITE INSPECTION WAS SCHEDULED FOR A LATER DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 786329 | O-ARM 1000 IMAGING SYSTEM | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00027-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |