FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 6135430 · Received November 30, 2016

Report

Report Number
2032227-2016-46703
Event Type
Malfunction
Date Received
November 30, 2016
Date of Event
October 17, 2016
Report Date
November 16, 2016
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH NO BUTTON RESPONSE DUE TO UNLOCKED KEYPAD CONNECTOR ON LCD BOARD. ANALYSIS WAS UNABLE TO PERFORM FUNCTIONAL CHECK INCLUDING REWIND AND BASIC OCCLUSION, OCCLUSION, PRIME, EXCESSIVE NO DELIVERY, DISPLACEMENT, SELF TEST, UNEXPECTED RESTART TEST AND ALL OPERATING CURRENT MEASUREMENT DUE TO NO BUTTON RESPONSE. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND STRIPPED BATTERY CAP COIN SLOT. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE UNABLE TO REMOVE THE BATTERY CAP OUT OF THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 360 MG/DL AT THE TIME OF INCIDENT. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL NEED TO BE REPLACED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
787258 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 67 YR