FDA Adverse Event
Other
Summary report: N
VIKING
MDR report key: 613477
·
Received June 6, 2005
Report
- Report Number
- 1225750-2005-00009
- Event Type
- Other
- Date Received
- June 6, 2005
- Date of Event
- May 6, 2005
- Report Date
- June 3, 2005
- Manufacturer
- LIKE AB
- Product Code
- FSA
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- WV, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
IN 2005, A PT FELL WHILE BEING RAISED IN A VIKING M PT LIFT/HOIST. ACCORDING TO THE FACILITY, THE PT WAS BEING LIFTED FROM A WHEELCHAIR WHEN THE SLING HANGER BAR'S ADAPTER BROKE. THIS CAUSED THE PT TO FALL APPROX 3 INCHES BACK INTO THE WHEELCHAIR. THE PT REPORTEDLY WAS NOT INJURED. THE EQUIPMENT WAS TAKEN OUT OF SERVICE UNTIL INSPECTED BY LIKO'S LOCAL DISTRIBUTOR THREE DAYS LATER. UPON INSPECTION IT WAS OBSERVED THAT THE SLING HANGER BAR'S ADAPTER HAD BEEN REMOVED. THIS UNAPPROVED COMPONENT REMOVAL LED TO UNSAFE SIDE LOAD FORCES UPON THE HANGER BAR'S CENTER BOLT, WHICH BROKE DURING THE LIFT. IT IS THE POSITION OF LIKO INC. THAT IF THE ADAPTER HAD NOT BEEN REMOVED THE EQUIPMENT WOULD NOT HAVE BROKEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIKING | PATINET LIFT/HOIST | FSA | LIKE AB | VIKING M | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |