O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Report
- Report Number
- 1723170-2016-01255
- Event Type
- Malfunction
- Date Received
- November 30, 2016
- Date of Event
- December 30, 2015
- Report Date
- December 21, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K092564
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- RADIOLOGIC TECHNOLOGIST
Narratives
THE SOFTWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS RELATED TO A SOFTWARE ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION SOFTWARE ANOMALY TRACKING DATABASE. A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE IMAGING SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. THE MEDTRONIC REPRESENTATIVE WAS UNABLE TO REPLICATE THE REPORTED ISSUE. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA (B)(6) DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 (B)(6). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
CORRECTION DEVICE UNIQUE DEVICE IDENTIFIER (UDI) NOW PROVIDED.
A SITE REPRESENTATIVE REPORTED THAT, WHILE IN A SPINAL FUSION PROCEDURE, THE DOOR FOR THE IMAGING SYSTEM WOULD NOT OPEN. THE SITE REPRESENTATIVE REPORTED THAT THE SYSTEM WAS AROUND THE PATIENT AT THE TIME. TO TROUBLESHOOT THE ISSUE, THE SYSTEM WAS RESTARTED WITHOUT RESOLUTION. THE SITE PRESSED THE DOOR OVERRIDE BUTTON ALONG WITH THE "M" BUTTON ON THE PENDANT, THIS ALLOWED THE DOOR TO OPEN AND RESOLVED THE ISSUE. THE SURGEON COMPLETED THE PROCEDURE WITH THE USE OF THE IMAGING SYSTEM. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF 10 MINUTES. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 786022 | O-ARM 1000 IMAGING SYSTEM 3RD EDITION | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00028-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |