DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2016-00847
- Event Type
- Malfunction
- Date Received
- November 29, 2016
- Date of Event
- November 4, 2016
- Report Date
- November 4, 2016
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
AN ISI TECHNICAL FIELD SPECIALIST (TFS) PERFORMED AN ONSITE INVESTIGATION AND REPLACED THE MTML AND RAC. INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE UNITS INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. THE INVESTIGATION RESULTS ARE AS FOLLOWS: THE REPORTED FAILURE COULD NOT BE REPRODUCED; THE RAC WAS INSTALLED INTO THE PRINTED CIRCUIT ASSEMBLY (PCA) TEST SYSTEM AND SINE CYCLE RAN FOR 10 MINUTES, TWO POWER CYCLES AND SAT IDLE FOR ONE HOUR, STILL NO TROUBLE WAS FOUND. THE REPORTED FAILURE WAS REPRODUCED DURING CALIBRATION OF THE MTML UNIT. VISUAL INSPECTION WAS PERFORMED AND THE OPTO BUTTON PADS AND GIMBAL GRIP PADS WERE FOUND TO BE WORN OUT. THE OPTO BUTTON PADS AND GIMBAL GRIP PADS WILL BE REPLACED. THIS COMPLAINT IS BEING REPORTED DUE TO A DA VINCI SYSTEM MALFUNCTION RENDERING THE DA VINCI SYSTEM UNAVAILABLE FOR USE AFTER THE START OF A SURGICAL PROCEDURE. ALTHOUGH NO PATIENT HARM OCCURRED, IF THIS MALFUNCTION WERE TO RECUR IT COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
DURING A DA VINCI ASSISTED SIGMOID COLECTOMY SURGICAL PROCEDURE, THE SITE CONTACTED INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT BECAUSE THEY WERE GETTING NON RECOVERABLE SYSTEM ERROR MESSAGES. THE SITE ATTEMPTED TO HARD POWER CYCLE THE SYSTEM; HOWEVER, THE NON RECOVERABLE SYSTEM ERROR MESSAGES CONTINUED TO OCCUR. WHEN ATTEMPTING TO RESET THE BREAKER ON THE SURGEON SIDE CONSOLE (SSC), THE SITE EXPERIENCED DIFFICULTY GETTING THE SSC TO POWER BACK ON. WHEN SYSTEM POWERED ON, THE SITE EXPERIENCED NON-RECOVERABLE FAULTS POINTING TO MASTER TOOL MANIPULATOR (MTM)LEFT AND THE REMOTE ARM CONTROLLER BOARDS (RAC). AT THIS TIME, THE SURGEON ELECTED TO CONVERT THE PROCEDURE TO A LAPAROSCOPIC SURGERY. NO PATIENT HARM WAS REPORTED AND THE PLANNED SURGICAL PROCEDURE WAS COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 783481 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A60P9B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |