FDA Adverse Event Summary report: N

65 SCOOP STRETCHER

MDR report key: 6132421 · Received November 29, 2016

Report

Report Number
1523574-2016-00033
Date Received
November 29, 2016
Date of Event
November 9, 2016
Report Date
January 30, 2017
Manufacturer
FERNO-WASHINGTON, INC.
Product Code
FPP
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT PROVIDED AN INCIDENT REPORT CLARIFYING THEY EXPERIENCED DIFFICULTY IN EXTENDING THE STRETCHER INTO POSITION AND THEY BELIEVED THERE TO BE SOME TYPE OF GRINDING WHEN OPENING AND CLOSING THE STRETCHER. THE SCOOP STRETCHER WAS RETURNED AND A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED. THE UNIT WAS LAID FLAT ON A TABLE AND WAS THEN CYCLED OPEN AND CLOSED WITH NO ISSUES. MULTIPLE SCRATCHES AND DENTS WERE OBSERVED ON THE STRETCHER. THE STRETCHER WAS ALSO REVIEWED FOR ANY SHARP EDGES; HOWEVER, NO SHARP EDGES WERE FOUND. THE REPORTED COMPLAINT COULD NOT BE DUPLICATED AND IT IS UNKNOWN HOW THE EMT ALLEGEDLY CUT HIS HAND ON THE STRETCHER. A FOLLOW UP COMMUNICATION WAS MADE TO THE EMT AND HE STATED THE LACERATIONS WERE SUPERFICIAL IN NATURE AND HE DID NOT SEEK ANY MEDICAL INTERVENTION.

Description of Event or Problem · 1

WE RECEIVED A REPORT FROM A DISTRIBUTOR THAT THEIR CUSTOMER WAS OPERATING THE SCOOP STRETCHER AND A MEDIC WAS ALLEGEDLY CUT ON A SHARP EDGE. IT WAS NOTED THERE WAS NO INJURY TO THE PATIENT. DETAILS ON THE ALLEGED INJURY AND ANY REQUIRED MEDICAL INTERVENTION WAS NOT PROVIDED IN THE INITIAL REPORT. THIS INCIDENT WILL BE FURTHER INVESTIGATED.

Description of Event or Problem · 1

WE RECEIVED A REPORT FROM A DISTRIBUTOR THAT THEIR CUSTOMER WAS OPERATING THE SCOOP STRETCHER AND A MEDIC WAS ALLEGEDLY CUT ON A SHARP EDGE. IT WAS NOTED THERE WAS NO INJURY TO THE PATIENT. DETAILS ON THE ALLEGED INJURY AND ANY REQUIRED MEDICAL INTERVENTION WAS NOT PROVIDED IN THE INITIAL REPORT. THIS INCIDENT WILL BE FURTHER INVESTIGATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
782489 65 SCOOP STRETCHER 65 SCOOP STRETCHER FPP FERNO-WASHINGTON, INC. 0108037

Patients

Seq Age Sex Outcome Treatment
1 66 YR