FDA Adverse Event Injury Summary report: N

EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE, 23MM

MDR report key: 6130637 · Received November 28, 2016

Report

Report Number
2015691-2016-03509
Event Type
Injury
Date Received
November 28, 2016
Date of Event
November 10, 2016
Report Date
November 10, 2016
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPT
PMA / PMN Number
P130009
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PER THE INSTRUCTIONS FOR USE, VALVE MIGRATION REQUIRING INTERVENTION IS A POTENTIAL ADVERSE EVENT ASSOCIATED WITH TRANSCATHETER AORTIC VALVE REPLACEMENT. ACCORDING TO LITERATURE REVIEW, VALVE MIGRATION RESULTS WHEN FORCES ACTING ON THE TRANSCATHETER HEART VALVE (THV) OVERCOME THE STRENGTH OF ATTACHMENT OF THE VALVE TO THE AORTIC WALL. STENT VALVES ARE SUBJECTED TO ANTEGRADE EJECTION FORCES DURING SYSTOLE. LESS-THAN-SEVERE AND NON-UNIFORMLY DISTRIBUTED CALCIFICATION OF THE NATIVE LEAFLETS, INCORRECT BIOPROSTHETIC VALVE SIZING, AND INCOMPLETE FRAME EXPANSION, CAN CONTRIBUTE TO VALVE MIGRATION. ADDITIONALLY, RESIDUAL OVERHANGING LEAFLETS CAN EXERT DOWNWARDS FORCE DURING DIASTOLE, CAUSING MIGRATION OF THE THV TOWARDS THE LEFT VENTRICLE. THE EDWARDS THV PATIENT SCREENING MANUAL ADVISES THE OPERATOR ON PRE-PROCEDURE ASSESSMENT OF THE AORTIC VALVE AND ROOT, TAKING INTO CONSIDERATION THE DEGREE AND DISTRIBUTION OF NATIVE LEAFLET CALCIFICATION. THE PROCEDURAL DIDACTIC INSTRUCTS THE OPERATOR ON PROPER POSITIONING AND DEPLOYMENT OF THE VALVE, INCLUDING ALL PROCEDURAL AND ANATOMICAL CONSIDERATIONS. PHYSICIANS ARE EXTENSIVELY TRAINED BY EDWARDS BEFORE THEY ARE QUALIFIED TO USE THE TRANSCATHETER HEART VALVE (THV). TRAINING INCLUDES PATIENT SCREENING, DEVICE PREPARATION, APPROACH, DEPLOYMENT, IMAGING, PROCEDURE-SPECIFIC TRAINING MANUALS AND PROCTORED PROCEDURES. THE CORRECT SIZING, ALIGNMENT AND POSITIONING OF THE DEVICE ARE EMPHASIZED AS KEY FACTORS TO THE PLACEMENT AND FIXATION OF THE DEVICE. IN THIS CASE, THE CAUSE FOR THE MIGRATION OF THE VALVE CANNOT BE DETERMINED; HOWEVER, THE VENTRICULAR MALPOSITION MAY HAVE CONTRIBUTED TO THE VALVE MIGRATION. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

AS REPORTED THROUGH OUR (B)(4) AFFILIATE, ONE MONTH AFTER A 23MM SAPIEN XT VALVE IMPLANT, FOLLOW UP ECHO REVEALED THAT THE VALVE HAD MIGRATED TOWARDS THE LVOT, WITH ELEVATED GRADIENTS AND THE PATIENT WAS SYMPTOMATIC (PRE-SYNCOPE). THE VALVE WAS EXPLANTED AND A SURGICAL VALVE WAS IMPLANTED. THE PATIENT IS DOING WELL. PER FOLLOW-UP WITH PHYSICIAN, THE THV SIZING WAS GOOD, MODERATE CALCIFICATION WAS PRESENT, AND THE VALVE MAY HAVE BEEN IMPLANTED LOWER THAN THE PHYSICIAN ORIGINALLY THOUGHT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
780283 EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE, 23MM AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES 9300TFX23

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention