PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2016-08363
- Event Type
- Injury
- Date Received
- November 28, 2016
- Date of Event
- October 25, 2016
- Report Date
- March 24, 2017
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). THE DEVICE WAS RETURNED FOR ANALYSIS AND A MARKER PATENCY TEST WAS PERFORMED, WHICH IDENTIFIED AN INABILITY TO FLUSH THE MARKER PORT. ABBOTT VASCULAR AV REVIEWED THE LOT HISTORY RECORD AND THERE WERE NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. THE COMPLAINT HANDLING DATABASE WAS REVIEWED AND IDENTIFIED NO EVENTS REPORTED FOR THE DEVICE OPERATES DIFFERENTLY THAN EXPECTED FROM THIS LOT. AV DETERMINED THAT THIS ISSUE APPEARS TO BE RELATED TO MANUFACTURING. ABBOTT WILL CONTINUE TO TREND THE PERFORMANCE OF THESE DEVICES.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER A PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY PROCEDURE. REPORTEDLY, AN UNKNOWN DEVICE FAILURE OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 779946 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 6031641 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |