ANIMAS VIBE
Report
- Report Number
- 2531779-2016-32237
- Event Type
- Injury
- Date Received
- November 28, 2016
- Report Date
- October 29, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- OYC
- UDI-DI
- 10840406100211
- PMA / PMN Number
- P130007
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PATIENT
Narratives
THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 02/22/2017. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 02/02/2017 WITH THE FOLLOWING FINDINGS: REVIEW OF THE PUMP DOWNLOAD SHOWED THE LAST BASAL DELIVERY AND THE LAST BOLUS DELIVERY WERE ON (B)(6) 2017. THE PUMP HISTORY SHOWED THE FOLLOWING MANUAL SUSPENDS AND MANUAL RESUMES ON (B)(6) 2016: AT 06:18 MANUAL SUSPEND AND MANUAL RESUME AT 06:19. AT 06:17 MANUAL SUSPEND AND MANUAL RESUME AT 06:18. AT 05:32 MANUAL SUSPEND AND MANUAL RESUME AT 06:17. THE PUMP PASSED DELIVERY ACCURACY TESTING AND WAS FOUND TO BE DELIVERING ACCURATELY AND WITHIN RANGE. THE TOTAL DAILY DOSES ADDED UP CORRECTLY TO REFLECT THE USER¿S PROGRAMMED BASAL RATES. NO ERRORS, ALARMS, WARNINGS OR DELIVERY INTERRUPTIONS OCCURRED DURING THE TESTING. INVESTIGATION DID NOT DUPLICATE THE COMPLAINT.
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE PATIENT WAS CONFUSED WHEN WAKENING AND THOUGHT THE PUMP WAS SUSPENDED; CHECKED ALARMS AND SUSPEND HISTORY, AND THEN SELF TREATED FOR A BLOOD SUGAR OF 60MG/DL. CUSTOMER SUPPORT ATTRIBUTED THE BG EXCURSION TO TRAINING/MISUSE. PATIENT RECEIVED NO TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE OF DIABETES CARE AND MANAGEMENT AND REMAINS ON PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 780114 | ANIMAS VIBE | INSULIN INFUSION PUMP | OYC | ANIMAS CORPORATION | 10840406100211 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Life Threatening |