FDA Adverse Event Injury Summary report: N

ANIMAS VIBE

MDR report key: 6128542 · Received November 28, 2016

Report

Report Number
2531779-2016-32237
Event Type
Injury
Date Received
November 28, 2016
Report Date
October 29, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
OYC
UDI-DI
10840406100211
PMA / PMN Number
P130007
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 02/22/2017. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 02/02/2017 WITH THE FOLLOWING FINDINGS: REVIEW OF THE PUMP DOWNLOAD SHOWED THE LAST BASAL DELIVERY AND THE LAST BOLUS DELIVERY WERE ON (B)(6) 2017. THE PUMP HISTORY SHOWED THE FOLLOWING MANUAL SUSPENDS AND MANUAL RESUMES ON (B)(6) 2016: AT 06:18 MANUAL SUSPEND AND MANUAL RESUME AT 06:19. AT 06:17 MANUAL SUSPEND AND MANUAL RESUME AT 06:18. AT 05:32 MANUAL SUSPEND AND MANUAL RESUME AT 06:17. THE PUMP PASSED DELIVERY ACCURACY TESTING AND WAS FOUND TO BE DELIVERING ACCURATELY AND WITHIN RANGE. THE TOTAL DAILY DOSES ADDED UP CORRECTLY TO REFLECT THE USER¿S PROGRAMMED BASAL RATES. NO ERRORS, ALARMS, WARNINGS OR DELIVERY INTERRUPTIONS OCCURRED DURING THE TESTING. INVESTIGATION DID NOT DUPLICATE THE COMPLAINT.

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE PATIENT WAS CONFUSED WHEN WAKENING AND THOUGHT THE PUMP WAS SUSPENDED; CHECKED ALARMS AND SUSPEND HISTORY, AND THEN SELF TREATED FOR A BLOOD SUGAR OF 60MG/DL. CUSTOMER SUPPORT ATTRIBUTED THE BG EXCURSION TO TRAINING/MISUSE. PATIENT RECEIVED NO TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE OF DIABETES CARE AND MANAGEMENT AND REMAINS ON PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
780114 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS CORPORATION 10840406100211

Patients

Seq Age Sex Outcome Treatment
1 63 YR Life Threatening