MDR report key: 6127045 · Received November 24, 2016

Report

Report Number
9617032-2016-00005
Event Type
Malfunction
Date Received
November 24, 2016
Date of Event
November 3, 2016
Report Date
January 6, 2017
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
NZU
PMA / PMN Number
K790366
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RESULTS: ONE USED SAMPLE AND THREE PHOTOS WERE RETURNED FOR EVALUATION. A VISUAL INSPECTION OF THE SAMPLE AND PHOTOS REVEALED A URINE CUP WITH THE EXTRA BENT AND DAMAGED NEEDLE INSIDE THE CUP. THE PHOTOS SHOWED THE POSITION OF THE EXTRA NEEDLE WHEN THE SECRETARY PRICKED HER FINGER ON IT. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF THE REPORTED LOT # 6190792. CONCLUSION: ALTHOUGH THE VISUAL INSPECTION CONFIRMED THE CUSTOMER'S INDICATED FAILURE MODE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED. OUR QUALITY ENGINEER NOTES THAT THE MOST LIKELY ROOT CAUSE WAS THAT THE EXTRA NEEDLE BECAME CAUGHT AND DAMAGED IN THE ASSEMBLY MACHINERY DUE TO TIMING ERROR ISSUES. BASED ON AN EVALUATION OF SEVERITY AND FREQUENCY IT WAS DETERMINED THAT NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR MONITORING OF CURRENT TRENDS.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A SECRETARY OBTAINED A CLEAN NEEDLE STICK INJURY FROM AN UNUSED BD VACUTAINER® PLASTIC URINE COLLECTION CUP. THE NEEDLE WAS BENT AND STICKING OUT FROM UNDER THE LABEL. THERE WAS NO REPORT OF MEDICAL INTERVENTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
777480 BD VACUTAINER® PLASTIC URINE COLLECTION CUP COLLECTION CUP NZU BECTON, DICKINSON AND COMPANY (BD) 6190792

Patients

Seq Age Sex Outcome Treatment
1 Other