ZOLL IVTM THERMOGARD XP
Report
- Report Number
- 3010617000-2016-00838
- Event Type
- Malfunction
- Date Received
- November 23, 2016
- Report Date
- November 23, 2016
- Manufacturer
- ZOLL CIRCULATION
- Product Code
- NCX
- UDI-DI
- 00849111075022
- PMA / PMN Number
- K072234
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THERMOGARD XP IVTM SYSTEM (S/N# (B)(4)) WAS SERVICED AT CUSTOMER SITE. HISTORICAL COMPLAINTS WERE REVIEWED FOR SERVICE INFORMATION RELATED TO THE REPORTED COMPLAINT AND THERE WAS NO PREVIOUS HISTORY OF COMPLAINT REPORTED FOR THERMOGARD XP IVTM SYSTEM S/N: (B)(4). A VISUAL INSPECTION WAS PERFORMED AND DRIP TRAY WAS MISSING (UNRELATED TO THE REPORTED COMPLAINT). A REVIEW OF THE EVENT LOG WAS PERFORMED AND FOUND NO ERRORS OR ANOMALIES. DURING FUNCTIONAL TESTING, ERROR MESSAGE TCMID: 02 (COOLING ENGINE DANFOSS ERROR) WAS NOTED, THUS CONFIRMING THE REPORTED COMPLAINT. IN SUMMARY, THE CUSTOMER'S REPORT OF THE THERMOGARD DISPLAYING TCMID:02 (COOLING ENGINE DANFOSS ERROR) WAS CONFIRMED DURING FUNCTIONAL TESTING. THE SYSTEM SOFTWARE WAS UPDATED. THE PIC16 BOARD AND DAN FOSS CONTROLLER WERE UPDATED TO RESOLVE THE PROBLEM. THE SYSTEM WAS CALIBRATED AND TESTED. THE CONSOLE PASSED FINAL FUNCTIONAL TEST WITH NO FURTHER ISSUE. THE THERMOGARD XP IVTM SYSTEM IS A REUSABLE DEVICE AND WAS MANUFACTURED IN 06/22/2010. THE DAMAGE FOUND IS A NORMAL WEAR AND TEAR FOR THE LIFE OF THE DEVICE.
IT WAS REPORTED THAT DURING PATIENT USE THE THERMOGARD XP IVTM SYSTEM (SN: (B)(4)) WAS DISPLAYING ERROR MESSAGE TCMID:02 (CE DANFOSS ERROR). THE CUSTOMER WAS ABLE TO CONTINUE AND COMPLETE THEIR TEMPERATURE MANAGEMENT THERAPY USING ANOTHER SYSTEM. THERE WERE NO REPORTS OF PATIENT CONSEQUENCES. PROBLEM OCCUR DATE IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 775729 | ZOLL IVTM THERMOGARD XP | THERMAL REGULATING SYSTEM | NCX | ZOLL CIRCULATION | 8700-0650 | 00849111075022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |