FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 6125640 · Received November 23, 2016

Report

Report Number
2032227-2016-44111
Event Type
Malfunction
Date Received
November 23, 2016
Date of Event
November 21, 2016
Report Date
January 11, 2017
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.(B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH THE OPERATING CURRENTS WITHIN SPECIFICATION AND PASSED THE SELF-TEST, UNEXPECTED RESTART ERROR TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. THE INSULIN PUMP PRIMED PROPERLY, NO EXCESSIVE NO DELIVERY ALARMS, AND NO DELIVERY ANOMALY OCCURRED DURING OUR TESTING. THE INSULIN PUMP HAD MINOR SCRATCHED LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP KEPT ALARMING NO DELIVERY. THE CUSTOMER HAS CHANGED THE INFUSION SET AND CHECKED TUBING. INSULIN WAS COMING THROUGH BUT IT KEPT ALARMING NO DELIVERY. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 7.2 MMOL/L. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
776793 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554LWWH

Patients

Seq Age Sex Outcome Treatment
1