FDA Adverse Event
Injury
Summary report: N
ADAC GENESYS
MDR report key: 61256
·
Received July 31, 1996
Report
- Report Number
- 61256
- Event Type
- Injury
- Date Received
- July 31, 1996
- Date of Event
- July 25, 1996
- Report Date
- July 30, 1996
- Manufacturer
- ADAC LABORATOERIES
- Product Code
- IYX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CT, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
HYPERCALCEMIC, TERMINAL LUNG PT IN NUCLEAR MEDICINE DEPT ON TABLE FOR A LUNG SCAN. PT'S LEFT ARM ACROSS CHEST FOR LATERAL VIEW WHILE CAMERA ROTATED ON CIRCULAR GANTRY. DUE TO WEAK, EMACIATED CONDITION, PT'S ARM SLIPPED OFF ABDOMEN AND WATCH BECAME CAUGHT ON THE COUNTERWEIGHT COVER WHERE ENCASED CABLE ENTERS THE COVER. AS CAMERA ROTATED, THE COVER PULLED THE WATCH CAUSING THE LEFT FOREARM TO BEND AND BREAK. EMERGENCY STOP BUTTON PUSHED. GANTRY WAS MOVED IN OPPOSITE DIRECTION TO RELEASE ARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADAC GENESYS | CAMERA-COUNTERWEIGHT | IYX | ADAC LABORATOERIES | 1357 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Required Intervention| S | CONTINUOUS HEPARIN INFUSION IN PERIPHERAL IV.| IN PERIPHERAL IV. |