FDA Adverse Event Injury Summary report: N

ADAC GENESYS

MDR report key: 61256 · Received July 31, 1996

Report

Report Number
61256
Event Type
Injury
Date Received
July 31, 1996
Date of Event
July 25, 1996
Report Date
July 30, 1996
Manufacturer
ADAC LABORATOERIES
Product Code
IYX
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CT, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

HYPERCALCEMIC, TERMINAL LUNG PT IN NUCLEAR MEDICINE DEPT ON TABLE FOR A LUNG SCAN. PT'S LEFT ARM ACROSS CHEST FOR LATERAL VIEW WHILE CAMERA ROTATED ON CIRCULAR GANTRY. DUE TO WEAK, EMACIATED CONDITION, PT'S ARM SLIPPED OFF ABDOMEN AND WATCH BECAME CAUGHT ON THE COUNTERWEIGHT COVER WHERE ENCASED CABLE ENTERS THE COVER. AS CAMERA ROTATED, THE COVER PULLED THE WATCH CAUSING THE LEFT FOREARM TO BEND AND BREAK. EMERGENCY STOP BUTTON PUSHED. GANTRY WAS MOVED IN OPPOSITE DIRECTION TO RELEASE ARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADAC GENESYS CAMERA-COUNTERWEIGHT IYX ADAC LABORATOERIES 1357 *

Patients

Seq Age Sex Outcome Treatment
1 44 YR Required Intervention| S CONTINUOUS HEPARIN INFUSION IN PERIPHERAL IV.| IN PERIPHERAL IV.