OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2016-03807
- Event Type
- Malfunction
- Date Received
- November 23, 2016
- Date of Event
- September 1, 2016
- Report Date
- October 28, 2016
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- UDI-DI
- 20385081120002
- PMA / PMN Number
- K122953
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PATIENT
Narratives
THE RETURNED PRODUCT WAS EVALUATED AND PERFORMED AS DESIGNED. NO KINK WAS OBSERVED IN THE CANNULA. NO MALFUNCTION OR OTHER PRODUCT CONDITIONS WERE FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA.
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE KINKED CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. OMNIPOD INSULIN MANAGEMENT SYSTEM ¿ USER GUIDE MODEL: UST400 14421-AW REV H / 2016 CHECKING YOUR BLOOD GLUCOSE 7 / PAGE 96. WARNING: TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER.
THE MOTHER REPORTED HER SON'S BLOOD GLUCOSE REACHED (300 MG/DL) AND THE CANNULA WAS KINKED. POD WORN BETWEEN 4 AND 24 HOURS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 775372 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 14000 | L42459 | 20385081120002 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 5 YR |