O-ARM 1000 IMAGING SYSTEM
Report
- Report Number
- 1723170-2016-03401
- Event Type
- Malfunction
- Date Received
- November 23, 2016
- Date of Event
- May 20, 2015
- Report Date
- November 23, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K050996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION AS NO PART WAS REPLACED. A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE OLD PATIENT DATABASE LOST. NOT ABLE TO REBUILD THE "OLD" DATABASE AS IT WAS NOT POSSIBLE TO REINITIALIZE OLD DATA. IMAGES ARE STILL ON THE SYSTEM BUT NOT VISIBLE. THE IMAGING SYSTEM PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. NO FURTHER ISSUE. THE SOFTWARE INVESTIGATION FOUND THAT A PROBABLE CAUSE WAS UNABLE TO BE DETERMINED WITHOUT FURTHER INFORMATION SINCE THE ON-GOING INVESTIGATION PROVED TO BE INCONCLUSIVE BASED ON THE INFORMATION PROVIDED. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA NEW ENGLAND DISTRICT OFFICE ON (B)(6) 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO (B)(4). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE (B)(6) 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
A SITE REPRESENTATIVE REPORTED THAT THE IMAGING SYSTEM WAS DISPLAYING AN ERROR MESSAGE: "WINDOWS COULD NOT START BECAUSE OF MISSING OR CORRUPT FILE." THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 774477 | O-ARM 1000 IMAGING SYSTEM | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00027-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |